Z-2093-2021: SoftGenomics version 4.1.15.6 SoftGenomics is a laboratory workflow and information management system to

Recall record Z-2093-2021 for Soft Computer Consultants, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2093-2021
Company in FDA record
Soft Computer Consultants, Inc.
FDA record creation date
Jul 19, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Software showing wrong results reported, PDF not match HIS.

Source facts

Action
The firm sent an " URGENT: MEDICAL DEVICE RECALL" letter dated 6/2/2021 to its sole consignee of the correction on 06/02/2021 by mail. The letter described the product, problem and actions to be taken. The consignee was instructed to do the following: Acknowledge receipt of this letter in the TMS task Give permission to load Hot Fix on Test environment Perform validation activities to completion Acknowledge Hot Fix has been moved to Live environment and is working as expected Type of action by Company: All clients on affected lines of software have been notified of the recall. SCC will provide Hot Fixes for all affected lines that will be delivered to all affected clients. Until the software correction is delivered, clients have been informed of an alternative workflow to avoid the problem. The firm provided information on a software update performed on 03/31/2021 and stated that, "Hot Fix has already been delivered." If you have any further questions, Please call 1-800-763-8522 (Technical Support 24/7).
Initiated
2021-03-31
Root cause
Software Design Change
Terminated
2022-04-15
Distribution
US Distribution to: CA only
Recall number
Z-2093-2021
Recall status
Terminated
CDRH record ID
187857
FDA record created
2021-07-19
Recall event number
88095

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗