Z-2093-2023: NeuMoDx SARS-CoV-2 Assay -IVD intended for the qualitative detection of SARS-CoV-2 RNA from nasal, nasop

Recall record Z-2093-2023 for NeuMoDx Molecular Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2093-2023
Company in FDA record
NeuMoDx Molecular Inc
FDA classification date
Jul 5, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

Potential for lot number discrepancies between the test strip packaging and the barcode label applied to the test strip that is read by the instrument for LOT 123909 of NeuMoDx SARS-CoV-2 Test Strips (REF 300800). The instrument registers the non-human readable text as lot number 123910 from the barcode label.

Source facts

Action
QIAGEN issued Urgent: Medical Device Correction Letter (PROM-22155-001) on 5/15/23 via email to Lab Director/Manager, Medical Director, Risk Manager,Safety Officer. Letter states reason for recall, health risk and action to take: If you have remaining stock of SARS-CoV-2 Test Strips LOT 123909 REF 300800, do not use it. Please contact QIAGEN Technical Service for a free-of-charge replacement. " Dispose of the product LOT 123909 in accordance with your national and local safety and environmental regulations. " Review this notice with your laboratory/medical director. Important: Forward this information to all individuals and departments within your organization using the above listed cartridges. If you are not the end user, please forward this notice to the product end user. " Complete Acknowledgement of Receipt attached to this letter as soon as possible. Assay strips from LOT 123909 can be identified from the box label, pouch label, and/or assay strip label. If you have any questions or concerns, please contact your local QIAGEN Technical Service Department through any of the following QIAGEN Subsidiaries: https://www.qiagen.com/us/contact-us/globalcontacts/subsidiaries
Initiated
2023-05-15
Root cause
Labeling mix-ups
Distribution
US Nationwide distribution in the states of FL, MI, NM, NY, TN, TX.
Recall number
Z-2093-2023
Recall status
Open, Classified
CDRH record ID
201145
FDA classified
2023-07-05
Recall event number
92490

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗