Open original FDA record ↗Recall record
- FDA identifier
- Z-2094-2015
- Company in FDA record
- Breas Medical AB
- FDA classification date
- Jul 16, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Unintended treatment termination could result from a keypad malfunction in some situations. The device erroneously interprets this as a Stop Treatment Instruction. An alarm will not sound, or be registered. Accessories and monitoring equipment connected to the Vivo 50 will stop functioning as the device enters stand-by mode.
Source facts
- Action
- Breas Medical sent Urgent Release: Field Safety Notice letters dated November 7, 2014 and Field Safety Notice dated November 21, 2014 to their customers. The letters identified the affected product, problem and the actions to be taken.
- Initiated
- 2014-11-07
- Root cause
- Device Design
- Terminated
- 2016-04-07
- Distribution
- Worldwide - United States Nationwide Distribution and the countries: Argentina, Belgium, Chile, Cyprus, Denmark, Ecuador, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Kuwait, Lebanon, Malaysia, Netherlands, Norway, Panama, Poland, Portugal, Russia, Saudi Arabia, Serbia, Singapore, Slovenia, Spain, Sweden, Switzerland, Thailand, Turkey, UAE and United Kingdom.
- Recall number
- Z-2094-2015
- Recall status
- Terminated
- CDRH record ID
- 131484
- FDA classified
- 2015-07-16
- Recall event number
- 69760