Open original FDA record ↗Recall record
- FDA identifier
- Z-2095-2021
- Company in FDA record
- Cardinal Health 200, LLC
- FDA record creation date
- Jul 19, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A caution statement is being added to remind the user to activate away from face and infant as reports of pack rupture or leaking during activation may resulting in the contents splashing into the user s eyes or onto the infant. Potential harms associated with contact of the liquid inside the pouch are rash, burn and inflammation.
Source facts
- Action
- Cardinal Health issued a "MEDICAL DEVICE PRODUCT ADVISORY" letter and Acknowledgement Form dated 6/9/2021 via overnight mail to the direct consignees on June 09, 2021. Letters states reason for issuance, health risk associated and action to take: The caution statement on the labeling will be updated to remind the user to activate away from face and infant. The full caution statement will now state: Caution: Activate away from face and infant. If pack ruptures or contents leak, flush contact area with water. Discard pack. If redness or irritation persists, consult a physician. What actions are needed from the Customer? Please review this enhanced caution statement with your staff to ensure product activation is performed away from the face and infant. For questions related to this notification and/or acknowledgement form that are not adequately addressed in this letter, please contact the market action team at: GMB-FieldCorrectiveAction@cardinalhealth.com or call 800-292-9332. In the event you have experienced quality problems or adverse events related to the products listed above, please contact GMB-CAH-Dist-Domestic@cardinalhealth.com
- Initiated
- 2021-06-09
- Root cause
- Package design/selection
- Terminated
- 2023-11-01
- Distribution
- US Nationwide Distribution to states of: AK, AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, IA ,ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WY.
- Recall number
- Z-2095-2021
- Recall status
- Terminated
- CDRH record ID
- 188044
- FDA record created
- 2021-07-19
- Recall event number
- 88173