Open original FDA record ↗Recall record
- FDA identifier
- Z-2096-2016
- Company in FDA record
- Stryker Sustainability Solutions
- FDA classification date
- Jul 22, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
Product defect; tip of two (2) Angiographic Catheters model # 10732203 broke off, prior to being used on the patient.
Source facts
- Action
- URGENT MEDICAL DEVICE RECALL RESTERILIZED ANGIODYNAMICS SOFT VU OMNI FLUSH ANGIOGRAPHIC CATHETER customer notification letters were sent 06/01/16. Customers were instructed to discontinue use of the recalled product immediately. Customers were told to complete the Recall Effectiveness Check Form and list the affected lots shipped to the facility. The form should be completed, signed, and returned to the local Stryker Sustainability Sales Representative, emailed to ssspfa@stryker.com, or mail to: Stryker Sustainability Solutions 1810 West Drake Drive Tempe, AZ 85283 Attn: Jodie Rueckert Customers will receive a credit for all affected devices returned.
- Initiated
- 2016-06-01
- Root cause
- Incorrect or no expiration date
- Terminated
- 2016-10-19
- Distribution
- Worldwide Distribution -- USA, to the states of TX, UT, PA, CA, VT, MA, WA, VA, NY, MD, and CT; and, the country of CANADA.
- Recall number
- Z-2096-2016
- Recall status
- Terminated
- CDRH record ID
- 147106
- FDA classified
- 2016-07-22
- Recall event number
- 74301