Skip to content

Public FDA record

Z-2096-2017: AMO 25GA Vitrectomy Cutter (Model Number NGP0025) and AMO 20GA Vitrectomy Cutter and Irrigation Sleeve (M

Recall record Z-2096-2017 for Abbott Medical Optics Inc. (AMO). View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2096-2017
Company in FDA record
Abbott Medical Optics Inc. (AMO)
FDA classification date
May 16, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Packaging error. 20GA Vitrectomy Cutters may be found in a 25GA package and 25GA Vitrectomy Cutters may be found in a 20GA package.

Source facts

Action
An Urgent Field Safety notice was sent on 3/30/17 to inform customers that AMO has voluntarily initiated this Action because a possibility exists that, due to a packaging error, 20GA Vitrectomy Cutters may be found in a 25GA package and 25GA Vitrectomy Cutters may be found in a 20GA package. Use of a Vitrectomy Cutter that is a different size than expected could lead to the need to alter the surgical technique, including conjunctival dissection, incision enlargement and scleral sutures.
Initiated
2017-03-30
Root cause
Packaging process control
Terminated
2017-08-23
Distribution
Worldwide distribution to SA, TR, AZ, NL, PT, SG, AU, JP, and HK.
Recall number
Z-2096-2017
Recall status
Terminated
CDRH record ID
154841
FDA classified
2017-05-16
Recall event number
77024

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗