Skip to content

Public FDA record

Z-2096-2021: AVSM3 SNF, Vital Signs Monitor The monitor is intended to be used to monitor noninvasive blood pressure

Recall record Z-2096-2021 for Mediana Co., Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2096-2021
Company in FDA record
Mediana Co., Ltd.
FDA record creation date
Jul 19, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

In some cases the equipment fails to power up and/or key or button unresponsive or not working.

Source facts

Action
The firm sent a "URGENT : Field Safety Notice(Correction)" email dated May 17, 2021 to its consignee. The email describes the product, problem and actions to be taken. The following actions were to be implemented: Implementation of action plan (1) Inventory stock a importer's hand: Mediana service engineer will update before ship to end users. (2) The products at end user's hand: Provide the modified new Software via USB memory stick and instruction for upgrading the S/W 3. Required actions for the user - INSPECT the inventory for the affected Model number, serial number(s), and date of manufacture. - IF USER HAVE affected devices Contact Customer Technical Services or Sales Representative for correction. - RETURN the enclosed acknowledgement form to Mediana USA service center. - UPGRADE THE S/W with the received s/w according to the instruction. - IF ANY Problem occurred during the upgrade Contact Customer Technical Services or Sales Representative. If you have any questions, contact QMS Representative at +82 70 7092 9945; email: don@mediana.co.kr. or Sales Representative at +82 70 7092 9525; email: chris.shin@mediana.co.kr. or Technical Services at 425-406-2262; email: Jason.kim@medianausa.com.
Initiated
2021-05-17
Root cause
Nonconforming Material/Component
Terminated
2022-11-04
Distribution
US Distribution to: Wisconsin only
Recall number
Z-2096-2021
Recall status
Terminated
CDRH record ID
188028
FDA record created
2021-07-19
Recall event number
88166

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗