Open original FDA record ↗Recall record
- FDA identifier
- Z-2096-2021
- Company in FDA record
- Mediana Co., Ltd.
- FDA record creation date
- Jul 19, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
In some cases the equipment fails to power up and/or key or button unresponsive or not working.
Source facts
- Action
- The firm sent a "URGENT : Field Safety Notice(Correction)" email dated May 17, 2021 to its consignee. The email describes the product, problem and actions to be taken. The following actions were to be implemented: Implementation of action plan (1) Inventory stock a importer's hand: Mediana service engineer will update before ship to end users. (2) The products at end user's hand: Provide the modified new Software via USB memory stick and instruction for upgrading the S/W 3. Required actions for the user - INSPECT the inventory for the affected Model number, serial number(s), and date of manufacture. - IF USER HAVE affected devices Contact Customer Technical Services or Sales Representative for correction. - RETURN the enclosed acknowledgement form to Mediana USA service center. - UPGRADE THE S/W with the received s/w according to the instruction. - IF ANY Problem occurred during the upgrade Contact Customer Technical Services or Sales Representative. If you have any questions, contact QMS Representative at +82 70 7092 9945; email: don@mediana.co.kr. or Sales Representative at +82 70 7092 9525; email: chris.shin@mediana.co.kr. or Technical Services at 425-406-2262; email: Jason.kim@medianausa.com.
- Initiated
- 2021-05-17
- Root cause
- Nonconforming Material/Component
- Terminated
- 2022-11-04
- Distribution
- US Distribution to: Wisconsin only
- Recall number
- Z-2096-2021
- Recall status
- Terminated
- CDRH record ID
- 188028
- FDA record created
- 2021-07-19
- Recall event number
- 88166