Z-2096-2025: Baxter SIGMA Spectrum Infusion Pump V8, Model Number 35700BAX2

Recall record Z-2096-2025 for Baxter Healthcare Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2096-2025
Company in FDA record
Baxter Healthcare Corporation
FDA classification date
Jul 17, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Certain Spectrum infusion pumps may have an incorrect version of software.

Source facts

Action
An URGENT MEDICAL DEVICE RECALL notification letter dated 6/20/25 was sent to customers. Actions to be Taken by Customers 1. Immediately locate Spectrum pumps with the affected serial numbers and remove them from service. The product code and serial number can be found on the bottom of the infusion pump. 2. Check and record the software version installed on the pump. The software on a V6 pump should begin with the number 6 and the software on a V8 pump should begin with the number 8. The software version is located on the startup screen upon powering up the pump as pictured on the next page. Additionally, the software version is indicated in the pump menu which can be accessed by following the steps. For V6: From the Pump Information screen, press the sw info soft key to display the software version screen. For V8: From the Select Care Area screen, press the options menu soft key. Select User Options and press OK. Select View Information and press OK. Select Pump Information and press OK to display the Pump Info screen. 3. Contact Baxter to confirm the software version on the affected pumps and to schedule service if required. Baxter Global Technical Services can be reached at 800-843-7867 (select option 1, then option 2), Monday through Friday, between 7:00 am and 7:00 pm Eastern Time. 4. If you received this communication directly from Baxter, acknowledge receipt of this notification by following the instructions on the enclosed reply instruction sheet, even if you have no remaining inventory. Acknowledging receipt will prevent you from receiving repeat notices. If you do not complete the acknowledgement, you will receive a phone call from OnProcess Technology on behalf of Baxter to confirm your receipt of this notification. 5. If you purchased this product from a distributor or wholesaler, please note that responding on the Baxter customer portal is not applicable. If a response is requested by your distributor or wholesaler, please respond to them
Initiated
2025-06-20
Root cause
Employee error
Distribution
US States: CA, FL, IL, KS, MI, MS, NY, OH, OK, and VA.
Recall number
Z-2096-2025
Recall status
Open, Classified
CDRH record ID
214421
FDA classified
2025-07-17
Recall event number
97072

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗