Open original FDA record ↗Recall record
- FDA identifier
- Z-2097-2015
- Company in FDA record
- Maquet Cardiovascular Us Sales, Llc
- FDA classification date
- Jul 17, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
There is a potential issue that could result to collision between the C-arm of the angiography system and the MAGNUS OR table system. After switching on the angiography-system, in rare cases, MAGNUS column provides wrong position data for the longitudinal movement leading to a collision.
Source facts
- Action
- Maquet sent a field correction letter dated June 8, 2015, via Fed Ex. Customers are advised of the issue with the device and provided instructions. Questions or additional information can be obtained from the local Maquet representative or Maquet Technical Support at 1-888-627-8383 (option 3 followed by option 1) Monday through Friday between the hours of 8:00 am and 6:00 pm EST.
- Initiated
- 2015-06-10
- Root cause
- Device Design
- Terminated
- 2016-04-18
- Distribution
- Worldwide Distribution - US Nationwide Distribution and to the countries of : France, Australia, Belgium, Brazil, Switzerland, Chile, China, Germany, Spain, United Kingdom, Hong Kong, Ireland, Israel, India, Italy, Japan, Kuwait, Kazakhstan, Mexico, Netherlands, Norway, Russia, Sweden, Turkey, Taiwan and South Africa.
- Recall number
- Z-2097-2015
- Recall status
- Terminated
- CDRH record ID
- 137828
- FDA classified
- 2015-07-17
- Recall event number
- 71446