A PBS (Pencil Beam Scanning) irradiation may pause for different reasons. After a pause, irradiation may restart from the beginning instead of recalculating the field based on the already delivered dose.
Source facts
Action
Target date for software update June 30, 2017.
A Field Safety Notice was issued on 4/10/2017 describing the problem and reminding users to follow guidance in the user's manual.
Initiated
2017-04-10
Root cause
Software design
Terminated
2018-02-16
Distribution
FL
Recall number
Z-2097-2017
Recall status
Terminated
CDRH record ID
154743
FDA classified
2017-05-03
Recall event number
76977