Skip to content

Public FDA record

Z-2097-2021: Illuminated Stepped Laser Probe Curved-for use in combination with 23G vitrectomy system Product Code: 75

Recall record Z-2097-2021 for Dutch Ophthalmic USA, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2097-2021
Company in FDA record
Dutch Ophthalmic USA, Inc.
FDA record creation date
Jul 20, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Barcode label extended expiration date. Barcode located at the short side of the outer box shows 2025-10-27 when scanned, while the correct expiry date is 2024-10-01

Source facts

Action
Dutch Ophthalmic issued Urgent Medical Device Recall letter on 4/14/21 via Certified Mail. Letter states reason for recall, health risk and action to take: Verify whether you have unused boxes of LOT 039406 in your inventory and quarantine product subject to this safety notice. " If returned contact Customer Service at Dutch Ophthalmic, USA at 1-800-753- 8824 x 1 to receive a Return Material Authorization (RMA) number and receive instructions on how to return unused boxes of the impacted product. " If not returning, verify that expiry dates in any hospital system are showing the correct expiry date of 2024- 10-01. If not, please ensure the expiry date is modified to reflect the actual expiry date. " For opened boxes of LOT 039406, verify that expiry dates in any hospital system are showing the correct expiry date of 2024-10-01. If not, please ensure the expiry date is modified to reflect the actual expiry date. Complete the attached Urgent Medical Device Safety Notice Reconciliation Form, sign it and return a scan or photo of the signed form by email to d.clement@dorcglobal.com. " Add the original completed and signed Urgent Medical Device Safety Notice Reconciliation Form referencing the Return Material Authorization number with the product to be returned. In the next days our sales representative will contact you to discuss replacement of the to be returned boxes and to verify whether there are additional questions or actions in which we can assist. If in the meantime you need any further information or support concerning this issue, please contact our local representative at Dutch Ophthalmic USA at 1-800-753-8824 x 1 for assistance.
Initiated
2021-04-14
Root cause
Process control
Distribution
Nationwide Distribution in the states of CA, FL, MA, and UT.
Recall number
Z-2097-2021
Recall status
Open, Classified
CDRH record ID
187983
FDA record created
2021-07-20
Recall event number
88146

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗