Skip to content

Public FDA record

Z-2098-2021: SoftCytogenetics

Recall record Z-2098-2021 for Soft Computer Consultants, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2098-2021
Company in FDA record
Soft Computer Consultants, Inc.
FDA record creation date
Jul 20, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Software is given wrong results going out to HIS on test that had reruns.

Source facts

Action
Dear SoftCytogentics Customer, The purpose of this letter is to advise you that SCC is voluntarily recalling SoftCytogenetics 4.1.11.20 and 4.1.12.11. There are software errors in these versions which SCC has designated to pose a potential risk-to-health issue. Reason for the Voluntary Recall: This task is to inform you of functionality in SoftCytogentics which SCC has designated to be a potential patient Risk-to-Health issue. For clients using Outbound Result Reporting with Resulting Profiles including Test Chapter macros, for cases that had multiple runs for reportable fields included to the Test Chapter macro, results from incorrect run may be sent outbound in HL7 message. Risk to Patient Care: When the results for test are processed with more than one run, and test uses the outbound reporting configuration of Resulting Profile* with Test Result macro added within, then some of results related to last reportable run can be sent outbound with values coming from any of the previous runs. These values are picked randomly from the previous runs. YOUR FACILITY DOES NOT CURRENTLY USE THIS CONFIGURATION. A hot fix would be required if you decide to implement this result reporting configuration prior to the next patch. Corrective Action: This issue will be corrected in the next software patch. Actions to be taken by the Customer/User: Please acknowledge receipt of this task and choose one of the following options: Workflow or functionality not used Agree to use alternative working solution Request hot fix if you decide to implement this result reporting configuration prior to the next patch.
Initiated
2021-03-31
Root cause
Software design (manufacturing process)
Terminated
2023-02-14
Distribution
US Nationwide Distribution in the states of CA, FL, MN and NJ.
Recall number
Z-2098-2021
Recall status
Terminated
CDRH record ID
187867
FDA record created
2021-07-20
Recall event number
88105

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗