Z-2098-2025: Varicam, Millennium VG, Millennium VG Hawkeye, Discovery VH Nuclear Medicine System

Recall record Z-2098-2025 for GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2098-2025
Company in FDA record
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING
FDA classification date
Jul 8, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Unintended radial detector motion may occur during patient setup or during patient scan if system does not have correct version of gantry software installed. Unintended detector motion may result in life-threatening injury.

Source facts

Action
Firm notified customers via Urgent Medical Device Correction letter dated June 20, 2025. Customers were instructed to check gantry software version. Instructions to do so are included in the letter. If system has the correct gantry software version, customers may continue to use their system--V.25BK for Infinia Running on Windows2000 (Win2K); C.15BK or C.1523 for Infinia on Windows XP; V.60BK on VG (all operating systems). If the software version is not the correct version, customers should discontinue use of system and contact GE HealthCare Service or local Service Representative. GE HealthCare will correct systems at no cost to the customer. Correction will include a hardware update and, if necessary, a software update. GE HealthCare representatives will contact customers to arrange for the correction. If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Initiated
2025-06-20
Root cause
Software design
Distribution
Worldwide - US Nationwide distribution.
Recall number
Z-2098-2025
Recall status
Open, Classified
CDRH record ID
214805
FDA classified
2025-07-08
Recall event number
97193

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗