Standard packed within NEFA kit has been assigned incorrectly. Calibrating with the incorrect value will cause a positive shift of up to +8% with both quality controls and patient samples
Source facts
Action
Randox notified the distribution center within the USA via email on 5/10/21.The distributor will then contact the customer directly.
Letter states reason for recall, health risk, and action to take:
Action to be taken:
" Please discard previous value sheets and ensure standard value is updated.
" Review results generated with the affected batches in line with the clinical profile of
the patient.
" Discuss the contents of this notice with your Medical Director
Initiated
2021-05-10
Root cause
Under Investigation by firm
Terminated
2021-11-10
Distribution
US Nationwide Distribution in the states of AR, CA, CT, NY, OH, PA, TN, and TX
Recall number
Z-2099-2021
Recall status
Terminated
CDRH record ID
187958
FDA record created
2021-07-20
Recall event number
88130