Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
FDA record creation date
Jul 20, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II
Reason stated in the source
Reveal LINQ with TruRhythm ICMs, that undergo a partial electrical reset appear to be programmed "ON", but are no longer able to detect and report Brady and Pause events to clinicians.
Source facts
Action
The firm began notifying their consignees on 06/02/2021 by letter. The letter explained the issue, provided patient management recommendations, and reported that a software update was anticipated in late 2021. For questions, contact your local Medtronic Representative or Medtronic Technical Services at 800-929-4043.
Initiated
2021-06-01
Root cause
Software design
Distribution
Worldwide Distribution
Recall number
Z-2102-2021
Recall status
Open, Classified