Skip to content

Public FDA record

Z-2103-2020: Kangaroo Epump Safety Screw Spike with Flush Bag, Anti-free Flow, 1000 mL, Product Code 775100

Recall record Z-2103-2020 for Cardinal Health Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2103-2020
Company in FDA record
Cardinal Health Inc.
FDA record creation date
May 22, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The feeding spike sets may leak at the interface of the tube and spike connector.

Source facts

Action
A Medical Device Product Advisory letter dated 12/23/19 was sent to customers. 1) INSPECT your inventory for affected product. 2) COMMUNICATE with all personnel that utilize the equipment the risk of leaks at the interface of the tube and spike connector. 3) INSPECT the feeding set for leaks after the pump has been primed and the infusion has started. " If a leak is detected the feeding set must be discarded and a new feeding set must be obtained. Inspect the new feeding set for leaks in the same manner. 4) NOTIFY any customers to whom you may have distributed or forwarded affected product about this product advisory. 5) RETURN the enclosed acknowledgment form via fax to 847-689-9101 or 614-652-9648 or email to gmb-fieldcorrectiveaction@cardinalhealth.com, whether or not you have affected product. Please contact the Customer Service group for any questions related to this action: " Hospital800-964-5227 " Federal Government800-444-1166 " Distributor800-635-6021 " All other customers888-444-5440 For questions related to the notification and/or acknowledgement form that are not adequately addressed in this letter, please contact the market action team at: GMB-FieldCorrectiveAction@cardinalhealth.com or call 800-292-9332.
Initiated
2019-12-26
Root cause
Process control
Distribution
Worldwide distribution.
Recall number
Z-2103-2020
Recall status
Open, Classified
CDRH record ID
178935
FDA record created
2020-05-22
Recall event number
84656

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗