Open original FDA record ↗Recall record
- FDA identifier
- Z-2104-2012
- Company in FDA record
- Intuitive Surgical, Inc.
- FDA classification date
- Jul 27, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
Specific lots of the Instrument Arm Drapes were manufactured with a sterile adaptor that may have difficulty engaging an instrument.
Source facts
- Action
- Intuitive Surgical sent an Urgent Medical Device Correction letters dated July 16, 2012 via Fed Ex to all affected customers in the US. The customers outside of the US were notified starting July 17, 2012. The letter identified the affected products, problem and actions to be taken. Customers were instructed to Identify affected lots, order replacement products and return both full and partial boxes of the affected products for credit as soon as unaffected products are available. All staff should be made aware of the issue. Customers were instructed to complete and return the acknowledgement form immediately. Questions and concerns should be directed to Intuitive Customer Service at 1-800-876-1310. EXPANSION: Intuitive discovered an additional lot (Lot DA121075) of Instrument Arm Drape IS2000, 20 pack that should have been included in the recall and expanded the recall on September 4, 2012. Letters to customers were sent by Fed Ex on September 5, 2012.
- Initiated
- 2012-07-13
- Root cause
- Nonconforming Material/Component
- Terminated
- 2013-06-12
- Distribution
- Worldwide Distribution - US Nationwide and the countries of: Australia, Spain, India, Italy and Saudi Arabia.
- Recall number
- Z-2104-2012
- Recall status
- Terminated
- CDRH record ID
- 110981
- FDA classified
- 2012-07-27
- Recall event number
- 62553