Skip to content

Public FDA record

Z-2106-2019: YXLON Cougar Cabinet X-Ray System Product Usage: The products are generally used for non-destructive tes

Recall record Z-2106-2019 for Comet Technologies Usa Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2106-2019
Company in FDA record
Comet Technologies Usa Inc
FDA record creation date
Aug 7, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Failure to meet the requirement of the performance standard for cabinet x-ray systems (21 CFR 1020.40) for a door interlock that causes physical disconnection of the power supply using only the movement of the door.

Source facts

Action
The firm sent a notification letter dated July 2019, to affected customers. The letter identified the affected products, problem and actions to be taken. Customers were informed that a service technician will be contacting you to schedule the retrofit service. For questions, call YXLON Service at (234) 284-7881.
Initiated
2018-10-01
Root cause
Radiation Control for Health and Safety Act
Terminated
2024-06-13
Distribution
US Nationwide Distribution
Recall number
Z-2106-2019
Recall status
Terminated
CDRH record ID
174622
FDA record created
2019-08-07
Recall event number
83295

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗