Z-2106-2020: Guardian Connect App CSS7200 iOS and Guardian Connect Transmitter GST4C used on the iPhone, iPad and iPo

Recall record Z-2106-2020 for Medtronic Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2106-2020
Company in FDA record
Medtronic Inc.
FDA record creation date
May 22, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Customers using the firm's continuous glucose monitoring system application on an iPhone, iPad or iPod Touch with ios software version 12, 12.1, or 12,2 are likely to experience a shortened transmitter battery life (approximately 4-5 days instead of the normal 7 days or more) after a full charge.

Source facts

Action
Urgent Medical Device Safety Notification Letters were sent to patients, health care providers (HCPs), distributors and clinical via email beginning on May 20, 2019 and via USPS certified mail beginning on May 22, 2019 notifying them of a potential issue with the transmitter battery life of the Connect Continuous Glucose Monitoring System. Customers using an Apple device with iOS version 12 or higher were advised to update to the latest version (3.2.4) of the Guardian Connect app on their iPhone as soon as possible. Customers not using iOS 12 or higher were encouraged to update to the latest version of the Guardian Connect app. Customers receiving emails were requested to submit email acknowledgements indicating they have read and understood the instructions provided in the Urgent Safety Notice Letter clicking on the button as evidence of receipt. Customers receiving paper-copy notifications were instructed to fill out a response card indicating they had read and understood the notice, and mail the card back with a pre-paid envelope. Customers with invalid email and/or physical addresses will be contacted by telephone. Customers with questions were encouraged to contact 24-Hour Technical Support Team at 1-800-646-4633.
Initiated
2019-05-20
Root cause
Software design
Distribution
US Nationwide distribution.
Recall number
Z-2106-2020
Recall status
Open, Classified
CDRH record ID
180063
FDA record created
2020-05-22
Recall event number
85091

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗