Z-2107-2023: REF: 218-1604, 1.6mm x 4mm Profile Zero Auto-Drive Screw, Rx Only, Sterile

Recall record Z-2107-2023 for Osteomed, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2107-2023
Company in FDA record
Osteomed, LLC
FDA classification date
Jul 7, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Screws for osteotomies have contain an undersize screwhead which has the potential to pass through the plate hole.

Source facts

Action
On June 6, 2023, OsteoMed/Acumed issued an "Urgent: Medical Device Recall" notification to affected consignees via FedEx Ground. OsteoMed/Acumed asked consignees to take the following actions: IMMEDIATELY stop using and/or distributing the product and complete the following actions: 1) Perform a physical count of your inventory. Identify any inventory of the lot number listed above that is still in its packaging. Locate and remove the affected, unopened, unused packages of screws, such that they are out of service (i.e., quarantined). 2) Record this data (product and quantity) on the survey found at the end of this letter (the Outstanding Field Action Survey). 3) Complete the remaining, applicable fields in the Outstanding Field Action Survey and either return it via email to Ellie.wood@acumed.net or a physical copy can be mailed to: OsteoMed/Acumed, Attn: Lauren Freytag- R23-002-TX, 3885 Arapaho Rd, Addison, TX 75001 4) If you have product to return, please contact Ellie.wood@acumed.net so that an appropriate return number and return instructions can be provided. 5) If you do not have product to return, you are still required to complete the form below. 6) For patients implanted with these affected devices, OsteoMed/Acumed advises physicians to monitor patients.
Initiated
2023-05-31
Root cause
Nonconforming Material/Component
Distribution
Worldwide - US Nationwide distribution in the states of AZ, CA, FL, GA, KS, KY, MD, MO, MT, NE, NY, TN, TX and the country of SG.
Recall number
Z-2107-2023
Recall status
Open, Classified
CDRH record ID
200997
FDA classified
2023-07-07
Recall event number
92479

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗