Z-2108-2020: ARTIS Pheno - Interventional Fluoroscopic X-Ray System Model: 10849000 - Product Usage: is a family of de

Recall record Z-2108-2020 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2108-2020
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA record creation date
May 22, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

If the C-arm leaves its intended travel path due to a fault within the drivetrain, movements may be impacted or impossible and the area of interest cannot be reached. In the event the C-arms movements are not possible, the system can only be returned to normal operation with the support of a field service engineer

Source facts

Action
Siemens issued Urgent Medical Device Correction dated 4/22/20 via AX074/19/S. Letter states reason for recall, health risk and action to take: Planned procedures may have to be terminated and performed on an alternative X-ray system. Please follow the standard emergency procedures you have in place in the event of system failures. It is recommended to consider reinforcing these emergency procedures until the update has been completed. Siemens will correct the error with a software patch in which the system will recognize a deviation of the intended travel path and unit movements are stopped immediately, similar to the activation of a proximity switch. The correction will be initiated via Update Instruction AX073/19/S. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Following the correction, the cause is eliminated, and any recurrence of this potential fault is prevented
Initiated
2020-04-22
Root cause
Software design
Terminated
2020-10-09
Distribution
US Nationwide distribution.
Recall number
Z-2108-2020
Recall status
Terminated
CDRH record ID
181337
FDA record created
2020-05-22
Recall event number
85608

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗