Open original FDA record ↗Recall record
- FDA identifier
- Z-2108-2020
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA record creation date
- May 22, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
If the C-arm leaves its intended travel path due to a fault within the drivetrain, movements may be impacted or impossible and the area of interest cannot be reached. In the event the C-arms movements are not possible, the system can only be returned to normal operation with the support of a field service engineer
Source facts
- Action
- Siemens issued Urgent Medical Device Correction dated 4/22/20 via AX074/19/S. Letter states reason for recall, health risk and action to take: Planned procedures may have to be terminated and performed on an alternative X-ray system. Please follow the standard emergency procedures you have in place in the event of system failures. It is recommended to consider reinforcing these emergency procedures until the update has been completed. Siemens will correct the error with a software patch in which the system will recognize a deviation of the intended travel path and unit movements are stopped immediately, similar to the activation of a proximity switch. The correction will be initiated via Update Instruction AX073/19/S. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Following the correction, the cause is eliminated, and any recurrence of this potential fault is prevented
- Initiated
- 2020-04-22
- Root cause
- Software design
- Terminated
- 2020-10-09
- Distribution
- US Nationwide distribution.
- Recall number
- Z-2108-2020
- Recall status
- Terminated
- CDRH record ID
- 181337
- FDA record created
- 2020-05-22
- Recall event number
- 85608