- FDA identifier
- Z-2109-2023
- Company in FDA record
- Baxter Healthcare Corporation
- FDA classification date
- Jul 7, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Potential for the device to crack where the Chest Base Prone support attaches to the carbon fiber operating room (OR) table rail.
Source facts
- Action
- Baxter Healthcare notified customers, via USPS letter, on 05/18/2023. Customers were instructed to 1. Identify affected product and discontinue use of the old design until a replacement unit is provided. 2. Contact Baxter Order Management team to arrange the return and replacement of affected devices. 3. Acknowledge receipt of this letter by completing a reply form on the customer portal, even if there is no inventory on hand, if received directly from Baxter. 4. Please return the acknowledgement form to your distributor/wholesaler according to their instructions. 5. If you distribute this product to other facilities or departments within your institution, please forward a copy of this communication to them. 6. If you are a dealer, wholesaler, distributor, or original equipment manufacturer (OEM) that distributed any affected product to other facilities, please notify your customers and check the associated box on the customer portal reply form. Baxter initiated an expansion for additional impacted units and consignees An updated customer notification was sent to all affected customers on 06/12/2023.
- Initiated
- 2023-05-18
- Root cause
- Device Design
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, Australia, China, Hong Kong, India, Japan, Korea, Malaysia, Singapore, Taiwan, Thailand, Belgium, Croatia, Denmark, Egypt, France, Ireland, Israel, Italy, Jordan, Kuwait, Lebanon, Netherlands, Norway, Poland, Qatar, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, Turkey, United Arab Emirates and United Kingdom.
- Recall number
- Z-2109-2023
- Recall status
- Open, Classified
- CDRH record ID
- 200570
- FDA classified
- 2023-07-07
- Recall event number
- 92375