Open original FDA record ↗Recall record
- FDA identifier
- Z-2110-2008
- Company in FDA record
- Acumed LLC
- FDA classification date
- Sep 20, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Product sterility may be compromised.
Source facts
- Action
- The firm issued four separate Urgent Notification: Product Removal were issued by the firm on 2-14-2008, 2-20-2008, 2-26-2008, and 3-25-2008 for the issue of poorly sealed sterile pouches. The notifications were reportedly sent by either FAX or e-mail. The firm explained the issue and requested customers to return the specified products. On 7/16/08, the firm issued a second recall notification to their customers. This letter informed the customer of the recall, listed the products affected, stated the reason for the recall, tells customers, "return existing inventories of the recalled products to Acumed... If you further distributed this product, please notify those customers/consignees of this recall. This recall is to be extended to the user level." The letter includes an effectiveness form to be filled out by the customer and returned to Acumed.
- Initiated
- 2008-02-14
- Root cause
- Packaging process control
- Terminated
- 2010-02-03
- Distribution
- Worldwide Distribution ---- USA including states of WA, NE, NC, UT, FL, CA, OR, TX, NY, IL, OH, MT, and Puerto Rico, and countries of United Kingdom, Australia, and Greece.
- Recall number
- Z-2110-2008
- Recall status
- Terminated
- CDRH record ID
- 71445
- FDA classified
- 2008-09-20
- Recall event number
- 48453