- FDA identifier
- Z-2110-2009
- Company in FDA record
- Physio Control, Inc.
- FDA classification date
- Sep 18, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Some LIFEPAK 15 Monitor/Defibrillators have a Non-Invasive Blood Pressure (NIBP) module that was supplied by CAS Medical Systems, Branford, Connecticut who is recalling the modules. The NIBP modules may be susceptible to sudden changes in cuff pressure. Sudden changes in cuff pressure may possibly be caused by bumping or squeezing the cuff. In some instances this may result in a screen display
Source facts
- Action
- The firm notified customers by letter dated August 2009. Customers will have the Noninvasive Blood Pressure (NIBP) module replaced. If additional questions about the notification, contact Technical Support by calling 1-800-442-1142, option 5, Monday-Friday 6:00 a.m. to 4:00 p.m. (Pacific Standard Time). .
- Initiated
- 2009-08-12
- Root cause
- Software design
- Terminated
- 2012-08-10
- Distribution
- Worldwide Distribution -- US, AUSTRALIA, GERMANY and NETHERLANDS.
- Recall number
- Z-2110-2009
- Recall status
- Terminated
- CDRH record ID
- 84613
- FDA classified
- 2009-09-18
- Recall event number
- 53087