Open original FDA record ↗Recall record
- FDA identifier
- Z-2111-2011
- Company in FDA record
- Alere North Americal
- FDA classification date
- Apr 28, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Potential for false negatives test results and negative/faint positive controls
Source facts
- Action
- Alere sent a Voluntary Product Recall letter dated March 21, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to indicate zero on the attached Recall Reply Form, sign, and fax to 1-877-877-0147, if they have no remaining product. For any remaining product customers were instructed to discontinue use and indicate the quantity of kits in the attached Recall Reply Form so that the firm can arrange for them to receive credit. Customers were instructed to destroy any affected units at their site. For any further distribution of product, customers will need to be contacted. Physicians and hospitals were advised to retest patients with negative results. Questions relating to this Voluntary Product Recall call 1-877-597-9586.
- Initiated
- 2011-03-21
- Root cause
- Other
- Terminated
- 2013-03-07
- Distribution
- Nationwide Distribution
- Recall number
- Z-2111-2011
- Recall status
- Terminated
- CDRH record ID
- 99308
- FDA classified
- 2011-04-28
- Recall event number
- 58407