Open original FDA record ↗Recall record
- FDA identifier
- Z-2111-2014
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Jul 23, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Siemens has become aware of the potential malfunction when using syngo Imaging XS, version VA70A or higher. In some cases a miscalculation may occur for the grey scale values in the functions Region of Interest, Pixel Lens, Edge enhancement and Histograms. This potential malfunction may occur for images with a pixel depth greater than 12 bit. All values calculated for images with a pixel d
Source facts
- Action
- Siemens sent a Customer letter dated June 30th, to all affected customers to inform them of a new software version VA70B_0114 that will be introduced to all syngo Imaging XS customers with version VA70A or higher to solve a potential issue regarding miscalculations done for the grey scale values in the functions "Region of Interest", "Pixel Lens", "Edge Enhancement", and "Histograms". For questions regarding this recall call 610-219-6300.
- Initiated
- 2014-06-30
- Root cause
- Software design
- Terminated
- 2015-02-24
- Distribution
- Nationwide Distribution
- Recall number
- Z-2111-2014
- Recall status
- Terminated
- CDRH record ID
- 128627
- FDA classified
- 2014-07-23
- Recall event number
- 68769