During Giraffe Shuttle charging, the batteries emit low concentrations of gases (outgassing) producing a strong sulfur odor.
Source facts
Action
GE Healthcare notified customers on 04/02/2020 via "Urgent Medical Device Correction" letter. The recall letter identified the affected product and serial numbers. The firm instructed the users to discontinue use of the product if at any time they smell a gas odor. GE Healthcare will correct all affected products at a later date when the correction is available at no cost.
Initiated
2020-04-02
Root cause
Device Design
Terminated
2024-05-06
Distribution
US Nationwide distribution in the states of AZ, CA, CO, DE, FL, GA, ID, IL, IN, LA, MD, MI, MS, NC, ND, NM, NV, NY, OH, Ok, OR, PA, SC, TN, TX, VA, WA, WI and WV.
Recall number
Z-2111-2020
Recall status
Terminated
CDRH record ID
181185
FDA classified
2020-05-22
Recall event number
85536