Open original FDA record ↗Recall record
- FDA identifier
- Z-2111-2021
- Company in FDA record
- Caption Health
- FDA record creation date
- Jul 22, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The firm is aware of a issue with ultrasound software that results in unintended video frames being included. This could lead to an incorrect automated ejection fraction (AutoEF) and image quality score (IQS) estimates.
Source facts
- Action
- On May 19, 2021 Caption health, Inc. contacted consignees via telephone to inform them about their recall. On May 20, 2021, the firm sent follow-up communication via via email. In addition to informing consignees about the recall, the firm also asked customers to take the following actions: 1. Until your system is updated, we recommend that all EF calculations are reviewed carefully using clinical expertise, to verify the accuracy of the results. 2. If you have any questions or concerns, please contact Caption health support line at 1-855-506-HELP between 7am-7pm ET. You can also email us at support@captionhealth.com at any time. We will follow-up with additional details relevant to your site by Monday May 24
- Initiated
- 2021-05-19
- Root cause
- Software design
- Terminated
- 2023-01-31
- Distribution
- US : AZ, DE, NY, IL, CT OUS: None
- Recall number
- Z-2111-2021
- Recall status
- Terminated
- CDRH record ID
- 187913
- FDA record created
- 2021-07-22
- Recall event number
- 88101