Z-2111-2023: Impella 2.5 intravascular micro axial blood pump, Product Number 005042

Recall record Z-2111-2023 for Abiomed, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2111-2023
Company in FDA record
Abiomed, Inc.
FDA classification date
Jul 14, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

There is a potential risk for unintentional interaction of the Impella motor housing with the distal stent of a transcatheter aortic valve replacement (TAVR) resulting in destruction of the impeller blades. This has resulted in low flow from the damaged Impella system. Systemic embolization of the fractured impeller material is a possibility.

Source facts

Action
An URGENT: MEDICAL DEVICE CORRECTION (NOTIFICATION) dated 6/14/23 was sent to customers. RECOMMENDATIONS: Clinicians are cautioned to position the Impella system carefully in patients with TAVR, and to be aware of this potential interaction. In this situation, clinicians should avoid repositioning while the device is spinning and should turn the device to P0 during repositioning or any movement that could bring the outlet windows into proximity to the valve stent structures. If there is low flow observed in a patient implanted with a TAVR while on Impella heart pump support, you should consider damage of the impeller and replace the Impella pump as soon as possible. ACTIONS TO BE TAKEN BY CUSTOMER/USER " Product is NOT being removed from the field and does not need to be returned. " Review, complete all fields, sign, and return the attached business response form (BRF) on the last page of this letter to the Recall Coordinator identified in this document. " Forward this notice to anyone in your facility that needs to be informed (i.e., those who manage, transport, store, stock, or use the subject products). " If any of the subject products have been forwarded to another facility, contact that facility and provide them with this notice. " Post a copy of this notice in a visible area for awareness of this field safety notice. " As with any medical device, adverse reactions or quality problems experienced with the use of this product should be reported to the FDA s MedWatch Adverse Event Reporting Program using the link below: https://www.fda.gov/safety/medwatch-fda-safety-information-and-adverse-event-reporting- program/reporting-serious-problems-fda At Abiomed, our priority is to our customers and their patients, and that includes the safe and effective use of our products. If you have questions or concerns regarding this notice, please contact (Shashi Thoutam) directly at +1(734) 262-6255 and/or your local clinical field staff. Thank you for your cooperation. Th
Initiated
2023-06-14
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide.
Recall number
Z-2111-2023
Recall status
Open, Classified
CDRH record ID
200239
FDA classified
2023-07-14
Recall event number
92321

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗