Z-2112-2018: Baxter SIGMA Spectrum Infusion Pump with Master Drug Library (Version 6), infusion pump, 35700BAX and 357

Recall record Z-2112-2018 for Baxter Healthcare Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2112-2018
Company in FDA record
Baxter Healthcare Corporation
FDA record creation date
Jun 7, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm will be updating the Instructions for Use for the product. Current IFUs instruct users to lower the primary bag 20 inches below the secondary bag using a hanger when performing a secondary infusion. However, the firm supplies 12.5-inch hangers with the Baxter disposable secondary sets.

Source facts

Action
On May 8, 2018, Baxter issued a Medical Device Correction letter to its customers. Customers were advised of the issue and instructed to do the following: 1. The SIGMA SPECTRUM Infusion Pumps can continue to be safely used with the Baxter-provided 12.5-inch hanger. 2. The updated Instructions for Use will be made available by May 31, 2018. At that time, you may access the updated IFU online at the Baxter Global Technical E-Service Center at URL: https://service.baxter.com/tsportal/#/login. For additional questions and concerns about this issue please contact Baxter Healthcare Medina at 800-356-3454, Option 1, between the hours of 7:00 am and 7:00 pm Eastern Time. 3. If you purchased this product directly from Baxter, complete the enclosed Baxter customer reply form and return it to Baxter by faxing it to 224-270-5457 or scanning and e-mailing it to fca@baxter.com, even if you do not have any inventory. Returning the Baxter customer reply form promptly will prevent you from receiving repeat notices. 4. If you purchased this product form a distributor, please note that the Baxter customer reply form is not applicable. If a reply form is provided by your distributor or wholesaler, please return it to the supplier according to their instructions. 5. If you distribute this product to other facilities or departments within your institution, please forward a copy of this communication to them. 6. If you are a dealer, wholesaler, distributor/reseller, or original equipment manufacturer (OEM) that distributed any affected product to other facilities, please notify your customers of this Medical Device Correction in accordance with your customary procedures.
Initiated
2018-05-08
Root cause
Other
Terminated
2019-04-08
Distribution
Worldwide Distribution - US Nationwide and the country of Canada.
Recall number
Z-2112-2018
Recall status
Terminated
CDRH record ID
164214
FDA record created
2018-06-07
Recall event number
80014

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗