Skip to content

Public FDA record

Z-2113-2011: Doctors Review System, OB-Windows/OB-View software release numbers 4.1 through 4.1.3, Digisonics, Inc., 3

Recall record Z-2113-2011 for Digisonics, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2113-2011
Company in FDA record
Digisonics, Inc
FDA classification date
Apr 29, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Study would on occasion have fetal heart rate and/or cervical length on the report that was not entered nor did it belong to that patient. Software displayed incorrect data for fetal heart rate and cervical length in OB ultrasound reports.

Source facts

Action
The firm, Digisonics Inc, notified the affected customers of software release via e-mail on January 24, 2007 and follow-up e-mail on February 7, 2007. The e-mails described the product, problem and the actions taken. The firm was able to trap the error and after this report the developer was able to determine the cause and a new executable was sent and installed at the customers site. If you have any question, please call 800-940-3240 or e-mail: service@digison.net.
Initiated
2007-02-09
Root cause
Other
Terminated
2011-05-02
Distribution
Worldwide distribution: USA (nationwide) and country of Canada.
Recall number
Z-2113-2011
Recall status
Terminated
CDRH record ID
99188
FDA classified
2011-04-29
Recall event number
58367

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗