Open original FDA record ↗Recall record
- FDA identifier
- Z-2113-2012
- Company in FDA record
- Sunrise Medical (US) LLC
- FDA classification date
- Jul 31, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
Reports of injury when the Padded Swing Away Armrest is used for full body weight. Owner's manual states that armrest is only designed for resting of forearm.
Source facts
- Action
- Sunrise Medical sent a Safety Notice letter dated June 25, 2012 to all affected customers. The letter identified the affected product, problem and actions to be taken. The letter request that clinicians/suppliers post this safety notice in a conspicuous location, making customers and therapists aware of the issue. The letter also encourage end users to read the owner's manual and follow the recommenations for proper usage of the recalled product. Customers are instructed to complete and return the Supplier Notification and Response Form. For questions contact Sunrise Medical Regulatory Affairs at 888-208-4901.
- Initiated
- 2012-06-26
- Root cause
- Labeling design
- Terminated
- 2012-11-01
- Distribution
- Worldwide Distribution - US (nationwide) and the countries of Argentina, Brazil, Columbia, Costa Rica, Mexico, Panama, Uruguay and Venezuela.
- Recall number
- Z-2113-2012
- Recall status
- Terminated
- CDRH record ID
- 110756
- FDA classified
- 2012-07-31
- Recall event number
- 62493