Open original FDA record ↗Recall record
- FDA identifier
- Z-2113-2013
- Company in FDA record
- DeRoyal Industries Inc
- FDA classification date
- Aug 28, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
DeRoyal Industries is recalling their product Insufflation Tubing Sets due to materials incompatibility which could possibly result in occlusion at or near the filter housing which could prevent the flow of gas through the device.
Source facts
- Action
- The firm notified customers via letters dated 7/15/2013 sent UPS and via DeRoyal Representatives. Customers were asked to notify DeRoyal of any remaining inventory to return for replacement product or credit. Non-responders will receive additional notifications. Effectiveness checks will be performed by contacting non-responders by phone.
- Initiated
- 2013-07-15
- Root cause
- Packaging process control
- Terminated
- 2015-05-20
- Distribution
- Worldwide Distribution-USA (nationwide) including DC and the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and the countries of Dubai, Arab Emirates, Quebec, Canada, Japan, and Ireland.
- Recall number
- Z-2113-2013
- Recall status
- Terminated
- CDRH record ID
- 120078
- FDA classified
- 2013-08-28
- Recall event number
- 65764