Skip to content

Public FDA record

Z-2113-2017: PARADYM Models: VR 8250, DR 8550, CRT-D 8750, RF VR 9250, RF DR 9550, RF CRT-D 9750, SonR 8770, RF SonR

Recall record Z-2113-2017 for Sorin Group Italia SRL - CRF. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2113-2017
Company in FDA record
Sorin Group Italia SRL - CRF
FDA record creation date
May 18, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Sorin is notifying physicians that certain defibrillators' battery depletion may be undetectable.

Source facts

Action
Consignees were notified on 4/11/17 via certified letter to make users aware of the issue and what actions to take to mitigate the potential risk of re-intervention associated with undetected battery depletion. Following and implanting physicians are being instructed to do one of two things: 1. Prior to lead revision: Deactivate the shock therapies to avoid further charging1, Wait 24 hours1, and Re-interrogate the ICD or CRT-D to check the updated battery status. If RRT is reached, initiate a device replacement. Or 2. If it is not possible to wait 24 hours prior to replacing the lead, the lead revision may be performed as scheduled and the device may be replaced prophylactically during the same procedure since the battery status is unknown. For questions contact your local LivaNova representave or call LivaNova Technical Services at 877-663-7674.
Initiated
2017-04-11
Root cause
Software design
Terminated
2020-05-22
Distribution
US Nationwide Distribution
Recall number
Z-2113-2017
Recall status
Terminated
CDRH record ID
155019
FDA record created
2017-05-18
Recall event number
77139

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗