Reason stated in the source
The firm will be updating the Instructions for Use for the product. Current IFUs instruct users to lower the primary bag 20 inches below the secondary bag using a hanger when performing a secondary infusion. However, the firm supplies 12.5-inch hangers with the Baxter disposable secondary sets.
Source facts
- Action
- On May 8, 2018, Baxter issued a Medical Device Correction letter to its customers. Customers were advised of the issue and instructed to do the following: 1. The SIGMA SPECTRUM Infusion Pumps can continue to be safely used with the Baxter-provided 12.5-inch hanger. 2. The updated Instructions for Use will be made available by May 31, 2018. At that time, you may access the updated IFU online at the Baxter Global Technical E-Service Center at URL: https://service.baxter.com/tsportal/#/login. For additional questions and concerns about this issue please contact Baxter Healthcare Medina at 800-356-3454, Option 1, between the hours of 7:00 am and 7:00 pm Eastern Time. 3. If you purchased this product directly from Baxter, complete the enclosed Baxter customer reply form and return it to Baxter by faxing it to 224-270-5457 or scanning and e-mailing it to fca@baxter.com, even if you do not have any inventory. Returning the Baxter customer reply form promptly will prevent you from receiving repeat notices. 4. If you purchased this product form a distributor, please note that the Baxter customer reply form is not applicable. If a reply form is provided by your distributor or wholesaler, please return it to the supplier according to their instructions. 5. If you distribute this product to other facilities or departments within your institution, please forward a copy of this communication to them. 6. If you are a dealer, wholesaler, distributor/reseller, or original equipment manufacturer (OEM) that distributed any affected product to other facilities, please notify your customers of this Medical Device Correction in accordance with your customary procedures.
- Initiated
- 2018-05-08
- Root cause
- Other
- Terminated
- 2019-04-08
- Distribution
- Worldwide Distribution - US Nationwide and the country of Canada.
- Recall number
- Z-2113-2018
- Recall status
- Terminated
- CDRH record ID
- 164215
- FDA record created
- 2018-06-07
- Recall event number
- 80014