Z-2113-2020: Cardiovascular Reprocessed ViewFlex Xtra ICE Catheter, REF D087031, OM St. Jude Medical, Size 9F (3.2mm),

Recall record Z-2113-2020 for Stryker Sustainability Solutions. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2113-2020
Company in FDA record
Stryker Sustainability Solutions
FDA classification date
Jun 1, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A subset of temporary intracardiac ultrasound catheters are labeled with the incorrect expiration date. They are mislabeled as having a shelf life of two years, while the actual shelf life is one year. The earliest affected product will expire in May 2020. Use of mislabeled devices may lead to hazards sterility breach, component breakdown or delay in treatment.

Source facts

Action
The firm will execute the recall in four phases: 1) Provide the Urgent Medical Device Recall Customer Notification Letter and Recall Business Reply Form to sales representatives via UPS on 3/27/2020 to notify affected customers; 2) Send the Urgent Medical Device Recall Customer Notification Letter and Recall Business Reply Form directly to affected customers; 3) Follow up phone calls and/or emails to nonrespondent customers as needed; 4) Additional follow up attempts to non-respondent customers as needed. The Urgent Medical Device Recall Customer Notification Letter and Recall Business Reply Form, dated March 2, 2020 instructs customers the following: The following Customer Actions are communicated in the Urgent Medical Device Recall Customer Notification Letter: Affected products may continue to be used prior to the one-year shelf life expiration. All products used prior to May 2020 are within the validated shelf life. 1. Please check your inventory for the affected devices. Affected devices may be identified through Serial Number (see Attachments 1 and 2). 2. Please complete the Urgent Medical Device Recall Business Reply Form (page 5) and indicate if any of the affected product remains in your inventory. Please return this completed form through one of the following methods: a. Email: SSSPFA@stryker.com b. To your Stryker Sustainability Solutions Sales Representative c. Fax: 480-763-5345 All faxes sent to 480-763-5345 will redirect to the SSSPFA@stryker.com email inbox. This email inbox will be monitored and customer responses will be documented to determine which customers have responded to the recall notification This form must be completed and returned to Stryker even if no affected product is found. 3. If any of the affected devices have been forwarded to additional facilities, please contact these facilities and communicate this recall to them. 4. Maintain awareness of this communication internally until all required ac
Initiated
2020-03-27
Root cause
Labeling False and Misleading
Terminated
2022-10-14
Distribution
US: AZ, CA, FL, GA, ID, IL, IN, KS, MA, MO, PA, TN, TX, VA, WA, WV OUS: None
Recall number
Z-2113-2020
Recall status
Terminated
CDRH record ID
180776
FDA classified
2020-06-01
Recall event number
85413

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗