Z-2114-2011: UniCel DxH 800 Coulter Cellular Analysis System Part Number: 629029 The UniCel DxH 800 Analyzer is a qu

Recall record Z-2114-2011 for Beckman Coulter Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2114-2011
Company in FDA record
Beckman Coulter Inc.
FDA classification date
Apr 29, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

For Panels containing CBC: The RBC and HCT are improperly corrected and not flagged when there is a WBC Pump Failure and the WBC ~ 140x103/uL. For Body Fluid Panels: The RBC is improperly corrected and not flagged when there is a WBC Pump Failure. IMPACT: Erroneous RBC and/or HCT could affect the interpretation and management of patients at risk with various types of anemias.

Source facts

Action
The recall communication was initiated on 10/20/2010, with Beckman Coulter forwarding a Product Corrective Action (PCA) letter with PCA Response Form (via US Postal Service for US customers) to all customers who purchased the UniCel DxH 800 Coulter Cellular Analysis System. This letter informs the customers about the problems identified and provides an action/resolution to this issue. When a WBC Pump Failure occurs, the WBC or TNC results will be flagged with 'R' and the WBC Pump Failure System message. The RBC and/or HCT results associated with these affected analyses also require review. This will be corrected in a future version of software. Beckman Coulter asked consignees to complete and return the enclosed response form within 10 days and share this information with laboratory staff and retain this notification as part of laboratory Quality System documentation. Questions were directed to Beckman Coulter Customer Service at 800-526-7694 in the United States or Canada, or consignees local Beckman Coulter representative.
Initiated
2010-10-20
Root cause
Other
Terminated
2014-09-02
Distribution
Nationwide in the US and worldwide to: Australia, Belgium, Canada, Colombia, Croatia, France, Germany, Greece, Hong Kong, India, Italy, Japan, Korea, Republic of, Macao, Malaysia, Myanmar, Netherlands, Norway, Philippines, Puerto Rico, Qatar, Russian Federation, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, Turkey and United Kingdom
Recall number
Z-2114-2011
Recall status
Terminated
CDRH record ID
98884
FDA classified
2011-04-29
Recall event number
58282

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗