Z-2114-2020: FlexLab Automation System used with the following Modules: High Volume Storage(P/N FLX-270, FLX-271, FLX-

Recall record Z-2114-2020 for Inpeco S.A.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2114-2020
Company in FDA record
Inpeco S.A.
FDA record creation date
May 26, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The sensors which activate the safety switches may be affected by delayed activation. In 1% of the cases, the response delay can be up to a maximum delay of 2 seconds, instead of 40ms expected by design.

Source facts

Action
Urgent: Medical Device Correction notification letters dated 3/30/20 were sent to customers. Action to be taken by the user After module protection removal, wait at least 2 seconds before accessing the impacted module (listed in Table 1). Your service provider will contact you to replace the sensor. Until the service visit please maintain awareness on this notice. Please complete and return the "Customer Letter Receipt Confirmation and Implementation Check form" attached to this letter within 15 days directly to the email address specified in the email communication. Please transfer this notice to whom it might concern. Contact reference person: For any clarification you may need, do not hesitate to contact: Eva Balzarotti - Regulatory Affairs Manager E-mail: Regulatorv.Affairs@inpeco.com Phone: (+41) 919118 224
Initiated
2020-04-07
Root cause
Component design/selection
Terminated
2024-06-13
Distribution
The products were distributed to the following US States: IL and NY.
Recall number
Z-2114-2020
Recall status
Terminated
CDRH record ID
181190
FDA record created
2020-05-26
Recall event number
85537

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗