Z-2115-2025: Access Vitamin B12 Reagent, Catalog Number 33000, Lot Number 439850

Recall record Z-2115-2025 for Beckman Coulter, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2115-2025
Company in FDA record
Beckman Coulter, Inc.
FDA classification date
Jul 11, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Beckman Coulter has received complaints of erroneous Vitamin B12 results when using Access Vitamin B12 Reagent lot 439850. The affected reagent lot may cause erroneously low or high Vitamin B12 results, leading to a delayed or improper diagnosis and/or repeat testing.

Source facts

Action
An URGENT MEDICAL DEVICE RECALL notification letter dated 2/19/25 was sent to customers. ACTION: Discontinue using Access Vitamin B12 reagent lot 439850. Discard all remaining reagent packs from this lot following your laboratory procedures. Beckman Coulter recommends sharing the content of this letter with your laboratory and/or medical director regarding the need to review previous patient results. Contact your local Beckman Coulter representative for replacement product requests. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product listed above to another laboratory, please provide them a copy of this letter. So that we are assured you have received this important communication, please respond within 10 days in one of the following ways: Electronically, if you received this communication via email. Manually, complete and return the enclosed Response Form. If you have any questions regarding this notice, please contact our Customer Support Center: From our website: http://www.beckmancoulter.com You may request replacement product by contacting your local Beckman Coulter Representative for replacement.
Initiated
2025-02-27
Root cause
Process control
Distribution
International distribution to the country of Andorra, Bulgaria, Denmark, Estonia, France, Germany, Greece, Ireland, Italy, Maldives, Netherlands, Portugal, Romania, Spain, Switzerland, Turkey, United Kingdom of Great Britain and Northern Ireland
Recall number
Z-2115-2025
Recall status
Completed
CDRH record ID
214222
FDA classified
2025-07-11
Recall event number
97018

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗