Open original FDA record ↗Recall record
- FDA identifier
- Z-2116-2010
- Company in FDA record
- Sonosite, Inc.
- FDA classification date
- Aug 2, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
PowerPark system has a potential for loss of power at the corresponding wall outlet and/or circuit due to poor alignment.
Source facts
- Action
- On 05/27/10, SonoSite began contacting their consignees by telephone. On 06/01/10, SonoSite sent out the URGENT-MEDICAL DEVICE RECALL letter to their consignees. The letter describes the recall product as SonoSite PowerPark. Customers should call the Technical Support at 1-877-657-8118 or e-mail to service@sonosite.com for any questions regarding the recall. The letter instructs consignees to stop using the PowerPark immediately and return the product back to SonoSite in exchange for a full refund. The consignees will be contacted by the firm's Sales Representative and/or Clinical Applications Specialist to schedule an appointment to retrieve the device and deliver a refund.
- Initiated
- 2010-05-27
- Root cause
- Nonconforming Material/Component
- Terminated
- 2011-04-18
- Distribution
- Worldwide Distribution -- USA, including states of GA, NM, PA, TX, VA, and WA and country of Australia.
- Recall number
- Z-2116-2010
- Recall status
- Terminated
- CDRH record ID
- 92115
- FDA classified
- 2010-08-02
- Recall event number
- 55874