Shock Hazard. The manufacturer failed to set the system power switch to "ON" during the line level leakage and high pot current draw testing.
Source facts
Action
Siemens Medical Solutions USA, Inc. sent a Customer Safety Advisory Notification letter out on May 15, 2009. Service engineers were to visit each customer to conduct necessary testing.
For questions regarding this recall call 650-694-5993.
Initiated
2009-05-15
Root cause
Release of Material/Component prior to receiving test results
Terminated
2011-04-29
Distribution
Worldwide Distribution - USA including CA, IL, MD, NC, OH, and UT and the countries of Australia, China, Denmark, Spain, South Korea, Norway, Poland, Russia and Sweden.
Recall number
Z-2116-2011
Recall status
Terminated
CDRH record ID
84460
FDA classified
2011-04-29
Recall event number
53013