Z-2116-2025: SwabCap Disinfecting Cap for Needlefree Connectors: SwabSleeve, REF: SCXT3-5-2000; SwabCap, 200-ct. Box,

Recall record Z-2116-2025 for ICU Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2116-2025
Company in FDA record
ICU Medical, Inc.
FDA classification date
Jul 11, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a manufacturing issue, disinfecting cap for needle-free connectors may have an incomplete seal between the foil lid and plastic container, which may result in isopropyl alcohol evaporation from the sponge, which may result in inadequate disinfection.

Source facts

Action
On 6/3/2025, correction notices were mailed and emailed to customers who were asked to do the following: 1) Inform all potential users of the product in your organization of this notification. 2) Discard the affected devices following your institution's process for discarding product. If discarding the device is not immediately possible at your facility, the device should be quarantined until disposal is possible. 3) Complete and return the Customer Response Form to icumedical2818@sedgwick.com 5. DISTRIBUTORS: If you have distributed affected products to your customers, please immediately forward this notice to them. Request that they complete the response form and return it to icumedical2818@sedgwick.com. For further inquiries, please contact firm at the following: Global Complaint Management, globalcomplaints@icumed.com,1-(866)-216-8806; Customer Service, customerservice@icumed.com, 1-(800)-258-5361; Field Action Processing, icumedical2818@sedgwick.com, 1-(866)-328-0114
Initiated
2025-06-03
Root cause
Process control
Distribution
Worldwide - US Nationwide distribution including in the states of TX, SD, ID, LA, IN, CO, OK, MO, MN, OR, AZ, NC, TN, VA, CA, IL, FL, MA, OH, MD, PA, NJ, NY, WA, GA, UT, MI, MS, AR, AL, NV, KS, IA, WV, HI, KY, NM and the country of CA.
Recall number
Z-2116-2025
Recall status
Open, Classified
CDRH record ID
214003
FDA classified
2025-07-11
Recall event number
96903

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗