Philips Healthcare received reports from the field that certain Ingenuity Core systems running software version 3.5.4 exhibited intermittent swirl-like ring artifacts that may appear on reconstructed images. A patient rescan may be required if the images cannot be used for interpretation due to the swirl-like artifact.
Source facts
Action
Philips Medical Systems (Cleveland) Inc. sent an Urgent Field Safety Notice letter to affected customers , via certified letter with receipts returned for tracking purposes. The letter identified the affected product, problem and actions to be taken. Philips Healthcare is implementing software update version 3.5.5 to correct the problem. For questions contact your local Philips representative or local Philips Healthcare office.
Initiated
2016-04-01
Root cause
Software design (manufacturing process)
Terminated
2018-06-26
Distribution
US Nationwide Distribution in the states of CA, FL, ID, KS, MI, NY,OH,TX
Recall number
Z-2117-2016
Recall status
Terminated
CDRH record ID
147530
FDA record created
2016-07-01
Recall event number
74427