Open original FDA record ↗Recall record
- FDA identifier
- Z-2118-2013
- Company in FDA record
- Ventana Medical Systems Inc
- FDA classification date
- Aug 28, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Ventana Medical System is initiating the recall of the BenchMark and DISCOVERY series instruments because there is a possibility that the heating element could separate from the underside of the heater cap and cause a non-staining or inconsistent staining event. Additionally, the system may not detect this anomaly and therefore it may not generate an error message for the slide position.
Source facts
- Action
- Ventana sent an Urgent Medical Device Correction Notification letter dated June 16, 2008 to all affected customers. The letter identifed the affected product, problem and actions to be taken. The letter instructed customers to carefully read and perform all of the actions outlined in the notification letter concerning potential safety issue. Customers with concerns regarding the staining results on any patient specimen are instructed to follow their internal Quality Procedures regarding the review of patient reports. For question contact your local Customer Service Center.
- Initiated
- 2008-06-12
- Root cause
- Device Design
- Terminated
- 2013-08-29
- Distribution
- Worldwide Distribution - USA Nationwide and the countries of Canada, Europe, Asia, Japan.
- Recall number
- Z-2118-2013
- Recall status
- Terminated
- CDRH record ID
- 73710
- FDA classified
- 2013-08-28
- Recall event number
- 49516