Skip to content

Public FDA record

Z-2118-2020: Clearview Total Uterine Manipulator (model UM750), The device is labeled in part, "ClearView(R) TOTAL, 7

Recall record Z-2118-2020 for Clinical Innovations, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2118-2020
Company in FDA record
Clinical Innovations, LLC
FDA record creation date
May 26, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

An incorrect component used during manufacturing. The uterine manipulator tip support does not have the "notch" required to attach the spacer and or colpotomy cup. This may result in a delay in surgery and additional surgical interventions to completed the procedure. There was a report of an injury.

Source facts

Action
On July 2, 2019, the firm sent an field Safety Notice letter to consignees. In addition to informing consignees about the recall they asked that the following actions be taken: 1. Please provide the field safety notice to your hospitals. If you have further distributed this product, please identify your customers and notify them at once of this communication and/or contact Clinical Innovations with the contact information s that we can follow-up with the owner of the device. 2. Please quarantine any remaining product in your hospitals and completed the response card attached to this filed safety notice. A company representative will contact you and arrange for the return of the product. 3. If you have any questions, please call Clinical Innovations at 1-8880268-6222, M-F, 8:00 Am 5:00 PM MT
Initiated
2019-05-30
Root cause
Process control
Terminated
2020-08-24
Distribution
International distribution to the country Japan.
Recall number
Z-2118-2020
Recall status
Terminated
CDRH record ID
180166
FDA record created
2020-05-26
Recall event number
85144

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗