A software anomaly, in that the Multileaf Collimator (MLC) message displayed on the 4DITC on versions 8.6 and below, does NOT interlock beam delivery.
Source facts
Action
Varian Medical Systems sent an Urgent Medical Device Correction Letter on February 28, 2011, to all affected customers.
Varian has created the procedure described in CTB-GE-579 to inform users on proper interfacing of the demountable BrainLAB m3 on the Varian Clinac. The procedure ( CTB-GE-579) is enclosed and posted to the customer site: http://www.MyVarian.com.
This design fix in 4DITC v10.1 or later will be provided to all affected sites.
If you require further clarification, please feel free to contact your local Varian Customer Support District or Regional Manager.
USA and Canada - 1-888-VARIANS (888-827-4265)
Europe - + 41 41 749 8844
Initiated
2011-02-28
Root cause
Software design
Terminated
2012-03-30
Distribution
Class 2 Recall - Worldwide Distribution -- USA including states of ; CA, NV, NJ, GA, IL, and TX and countries of Australasia, Belgium, Brazil, China P.R., Denmark, France, Germany, Germany ( BrachyTherapy) Hong Kong, India Pvt. Ltd, Italy, Japan, Middle East, Netherlands, Russia, Spain, Switzerland and United Kingdom.
Recall number
Z-2119-2011
Recall status
Terminated
CDRH record ID
98761
FDA classified
2011-04-29
Recall event number
58230