Z-2119-2011: The 4D Integrated Treatment Console; Model Number: H51; Reference: 4DITC versions 8.6 and below; File Ref

Recall record Z-2119-2011 for Varian Medical Systems, Inc. Oncology Systems. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2119-2011
Company in FDA record
Varian Medical Systems, Inc. Oncology Systems
FDA classification date
Apr 29, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

A software anomaly, in that the Multileaf Collimator (MLC) message displayed on the 4DITC on versions 8.6 and below, does NOT interlock beam delivery.

Source facts

Action
Varian Medical Systems sent an Urgent Medical Device Correction Letter on February 28, 2011, to all affected customers. Varian has created the procedure described in CTB-GE-579 to inform users on proper interfacing of the demountable BrainLAB m3 on the Varian Clinac. The procedure ( CTB-GE-579) is enclosed and posted to the customer site: http://www.MyVarian.com. This design fix in 4DITC v10.1 or later will be provided to all affected sites. If you require further clarification, please feel free to contact your local Varian Customer Support District or Regional Manager. USA and Canada - 1-888-VARIANS (888-827-4265) Europe - + 41 41 749 8844
Initiated
2011-02-28
Root cause
Software design
Terminated
2012-03-30
Distribution
Class 2 Recall - Worldwide Distribution -- USA including states of ; CA, NV, NJ, GA, IL, and TX and countries of Australasia, Belgium, Brazil, China P.R., Denmark, France, Germany, Germany ( BrachyTherapy) Hong Kong, India Pvt. Ltd, Italy, Japan, Middle East, Netherlands, Russia, Spain, Switzerland and United Kingdom.
Recall number
Z-2119-2011
Recall status
Terminated
CDRH record ID
98761
FDA classified
2011-04-29
Recall event number
58230

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗