Skip to content

Public FDA record

Z-2119-2017: 10-1398M - SpF-PLUS Implantable Stimulator. Indicated as a spinal fusion adjunct to increase the probabi

Recall record Z-2119-2017 for EBI Patient Care, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2119-2017
Company in FDA record
EBI Patient Care, Inc.
FDA classification date
May 26, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Zimmer Biomet is conducting a medical device field action for the SpF PLUS-Mini and SpF XL IIB Implantable Spinal Fusion Stimulator products due to higher than allowed cytotoxicity levels found during routine process monitoring with the product.

Source facts

Action
An Urgent Medical Device Recall letter, dated April 20, 2017, was sent to the firm's consignees. The letter identified the affected product and the reason for the recall, as well as the risks involved. Consignees were instructed to quarantine affected product in their inventory and complete the Certification of Acknowledgement portion and the Inventory Return Certification Form in the letter. Affected product must be returned immediately along with the completed Inventory Return Certification Form. Questions or concerns should be directed to the customer call center at 1-800-447-3625 or via email to corporatequality.postmarket@zimmerbiomet.com.
Initiated
2017-04-20
Root cause
Other
Terminated
2018-10-17
Distribution
Nationwide distribution only.
Recall number
Z-2119-2017
Recall status
Terminated
CDRH record ID
155075
FDA classified
2017-05-26
Recall event number
77089

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗