Open original FDA record ↗Recall record
- FDA identifier
- Z-2119-2025
- Company in FDA record
- Waismed Ltd.
- FDA classification date
- Jul 17, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to manufacturing error, Intraosseous device built-in stabilizer may not be released properly after deployment and insertion.
Source facts
- Action
- On June 05, 2025, WaisMed issued an "Urgent: Medical Device Recall" notification to affected consignees via E-Mail. WaisMed asked consignees to take the follow actions: 1. Healthcare providers should immediately discontinue use of affected devices and consider alternative intraosseous access products to ensure continuity of patient care without compromising treatment timelines in critical situations. 2. Immediately discontinue, quarantine and remove all NIO A and NIO+ devices from the affected lots from all points of use, including emergency kits, crash carts, ambulances, and clinical storage areas. 3. Verify your inventory against the specific lot numbers listed in this recall notice to identify all affected products. 4. Share and post this Product Recall letter within your facility network. 5. Forward to any customers you may have distributed the product to ensure awareness. 6. Ensure the contents of this Product Recall are read and understood by those within your organization. 7. Please complete the attached Recall Acknowledgement and Receipt Form and return it to: RECALLS@SAFEGUARDMEDICAL.COM or mail to: Waismed Ltd. 10 Amal St. Afek Industrial Park, Rosh Ha'yin 4809234, Israel. Any questions directly associated with this recall should be directed to the: Recall Coordinator or International Director Quality and Regulatory Affairs via email: vigilanceil@safeguardmedical.com, phone: +972-9-9517444 or website: safeguardmedial.com.
- Initiated
- 2025-06-05
- Root cause
- Process change control
- Distribution
- Worldwide distribution: US (nationwide) to state of: NC and OUS (international) countries of: Slovenia, Australia, Brazil, Estonia, South Africa, U. Arab Emirate, Greece, Panama, Singapore, Italy, Germany, Portugal, Spain, Ukraine, Lithuania, Chile, Poland, Germany, New Zealand, Montenegro, Switzerland, Argentina, Poland, Austria, Sweden, United Kingdom, Italy, France, Romania, Netherlands, Croatia, Lithuania, New Zealand, Japan, Cyprus, and Israel.
- Recall number
- Z-2119-2025
- Recall status
- Open, Classified
- CDRH record ID
- 214388
- FDA classified
- 2025-07-17
- Recall event number
- 97061