Pumps submitted for service may not have been properly serviced before being returned to the customer.
Source facts
Action
Baxter visited the direct accounts starting on 3/24/11 informing them that the affected Flo-Gard pumps may have been released from the preventative maintenance with incorrect accuracy testing. The affected pumps were located, taken out of service and placed in quarantine on the same date. A technician not involved with the previous testing visited both accounts on 3/25/11 and performed preventative maintenance service of the three affected devices, and were found to be within specification.
Initiated
2011-03-24
Root cause
Employee error
Terminated
2011-04-29
Distribution
Puerto Rico
Recall number
Z-2122-2011
Recall status
Terminated
CDRH record ID
99440
FDA classified
2011-04-29
Recall event number
58482