Z-2122-2017: CSI, Orbital Atherectomy System (OAS) Saline Infusion Pump, Rx only, Part Number: 7-10014, Model Number:

Recall record Z-2122-2017 for Cardiovascular Systems Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2122-2017
Company in FDA record
Cardiovascular Systems Inc
FDA record creation date
May 20, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Cardiovascular Systems, Inc. (CSI). has initiated a recall to remove 1,396 7-10014 Saline Infusion Pumps (SIP) which were distributed by CSI between 07 April 2015 and 04 April 2017. The 7-10014 pumps may switch to stand-by during use requiring the pump to be reset prior to continuing treatment. To date, none of the associated complaints have resulted in any patient harm.

Source facts

Action
Consignees were sent on 4/14/2017 a CSI "Urgent Medical Device Recall" letter dated 14 APRIL 2017. The letter listed the Affected Product, Recall Description, SIP Replacement and Return & contact information. Consignees were notified the pumps will be replaced and the process is anticipated to be completed 31 August 2017. Consignees may continue to use their pumps however if a yellow light fault is observed and the troubleshooting per the IFU does not resolve the fault, discontinue use of the pump and notify CSI. Consignees were requested to complete and return the Customer Acknowledgement Form. For further information contact CSI sales Representative of Recall coordinator Jake Mellem at jmellem@csi360.com, 651-259-2819 - Tel.
Initiated
2017-04-14
Root cause
Device Design
Terminated
2017-10-02
Distribution
US: AK, AL, AR, AZ. , CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, AUSTRIA, GERMANY, JAPAN, SWITZERLAND.
Recall number
Z-2122-2017
Recall status
Terminated
CDRH record ID
154675
FDA record created
2017-05-20
Recall event number
76953

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗